pH - Diagnostics Menu - Calibration Data 4.16

pH - Diagnostics Menu - Calibration Data 4.16

Page 35

DOWN

DOWN

DOWN

DOWN

DOWN

DOWN

DOWN

DOWN

DOWN

DOWN

DOWN

DOWN

The Calibration Data menu is a series of read only screens which
allow the user to view the data collected during the last calibration.

RUN MODE

Press to view the first
Calibration Data screen.

UP

NOTE

PRESS THE AND KEYS

TOGETHER TO GO IMMEDIATELY BACK TO

RUN MODE

DOWN

1st POINT PH = 7.00
INPUT FROM PROBE = - 19 mV

Calculated efficiency based on 2
point calibration = 91 %

Calculated slope based on 2
point calibration = 59.16 mV/pH

Probe Temperature at which
calibration was performed = 19.4°C

2nd POINT PH = 4.00
INPUT FROM PROBE = + 143 mV

Calibration Mode
2 point calibration

Press to view the next
Calibration Data screen.

EFF 91%

>

59.16MV/PH

>

CALIBRATION DATA
CAL TEMP 19.4C >

CALIBRATION DATA
2P 4.00/ 143

>

CALIBRATION DATA
1P 7.00/ -19

>

MODE 2PT

>

CALIBRATION DATA

7.15pH 25.0C