Ph - led display menu - ph auto calibrate 4.23, Auto calibrate – Metex Shark Multi-Parameter Controllers & Analyzers User Manual

Page 47

Advertising
background image

S

H

A

R

K

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

pH - LED Display Menu - pH Auto Calibrate 4.23

pH - LED Display Menu - pH Auto Calibrate 4.23

Page 45

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

ENTER

AUTO CALIBRATE

CAL

Press the key to go
immediately back to Run mode

CAL

ENTER

ENTER

ENTER

ENTER

CAL

ENTER

ENTER

ENTER

ENTER

CAL

RUN

TO MANUAL CALIBRATE

TO MANUAL CALIBRATE

ENTER

The Controller will display bUF2, prompting
the user to place the probe in the second
buffer solution.

Be sure to clean and rinse the Probe first with
D.I. water and then insert it in the 4.00/10.00
buffer.
Press the key.

The Controller will display EFF for 3 seconds
to indicate it is calculating the probe efficiency
value.

It will then display the actual efficiency for a
few seconds and then revert back to run
mode.

Press the key again to confirm the entry.

If the displayed value is correct, press
the key.

The controller will flash "bUF2" for about 10
seconds while it reads the buffer. The
controller will determine which one of the
standard buffers is being used, and will set
the display to that value.

If the controller detects
an error, it will flash the
Err (Error) screen and
return back to Auto
Calibrate

The display will flash

Press the
key to select Auto
Calibration

The controller will flash "bUF1" for about 10
seconds while it reads the buffer. The
controller will determine which one of the
standard buffers is being used, either 4, 7 or
10, and will set the display to that value.

The Controller will display bUF1, prompting
the user to place the probe in the first buffer
solution.

Be sure to clean and rinse the Probe first with
D.I. water and then insert it in the 7.00 buffer.
Press the key.

If the displayed value is correct, press
the key.

Press the key again to confirm
the entry.

If the controller detects
a problem, it will flash
the Err (Error) screen
and return back to Auto
Calibrate

The display will flash

AUTO CALIBRATE
Auto Calibrate will
automatically detect the
buffers used to calibrate,
and set the display to the
appropriate settings.

MANUAL CALIBRATE
This menu allows the
user to manually
calibrate the probe, the
user enters the known
buffer values

Press the
key to select
Manual Calibration