Section 3 - Electrical Connections and Setup

3.14 Fuse Replacement

FUSE

P6

54

P1

P4

P3

4

2 3

1

7 8 9

5 6

15

13

11 12 14

10

17

16

24
23

21

22

29
28

26
25

27

63

64

62

67

66

65

57

56

55

61

60

59

58

31

30

52

53

51

Figure 3.8 Fuse Location

3.13 Service

SHARK SERVICE TO BE PERFORMED BY QUALIFIED PERSONNEL ONLY.

WARNING:

DISCONNECT LINE POWER TO THE UNIT

TO AVOID THE POSSIBILITY

OF ELECTRICAL SHOCK.

1. Proceed

after disconnecting line

power from the instrument.

2. Open the front panel by rotating the
quarter-turn fasteners, using a flat blade
screwdriver, to expose the relay board.

3. The fuse, F1, is located in the middle
of the relay board, directly above the
three terminal connectors.

4. Remove the open fuse and replace it
only with a fuse of the same type and
rating. REFER TO THE FUSE RATING
TABLES BELOW.

5. Close the front panel and secure
using the quarter-turn fasteners.

6. Restore power to the unit.

Fuse Rating Table for

120 volt operation

Fuse Rating Table for

240 volt operation

Fuse Type: Slo-Blo
fuse 5 x 20mm

Fuse Ratings: 250
VAC, 100mA

Fuse Type: Slo-Blo
fuse 5 x 20mm

Fuse Ratings: 250
VAC, 50mA

Dwg# N104-40