Conductivity - LED Display Menu - Conductivity
Manual Calibrate 6.23

Conductivity - LED Display Menu - Conductivity Manual Calibrate 6.23

Page 103

CAL

DOWN

ENTER

ENTER

ENTER

ENTER

UP

DOWN

UP

ENTER

ENTER

CAL

RUN MODE

MANUAL CALIBRATE

ENTER

ENTER

The Controller will display bUFF, prompting
the user to place the probe in the buffer
solution.
Be sure to clean and rinse the Probe first
with D.I. water and then insert it in the
Buffer.
Press the key.

Use the and keys to adjust
the display until it reads the correct value of
the buffer.

Press the key to accept the setting.

Press the key again to confirm
the change in calibration value.

The controller will flash "bUFF" for about
10 seconds while is reads the buffer. It
will then display the conductivity value.

The display will flash

MANUAL CALIBRATE
This menu allows the
user to manually
calibrate the probe, the
user enters the known
conductivity buffer value

Press the
key to select
Manual Calibration