Chapter 4 _________________________________________________________________ Operation

VAISALA ________________________________________________________________________ 59

1-point Adjustment of RH Measurement

The 1-point adjustment adjusts both offset and gain depending on the

adjustment condition. The same type of adjustment is done when turning

the RH trimmer.

Place the transmitter in the reference condition and allow it to stabilize

before entering the adjustment.

CRH [ONE] [x]<cr>

where
x = The reference humidity condition (%RH) that the transmitter

should be showing.

Example:

>pass 9000

>crh one 11

OK

2-point Adjustment of RH Measurement

CRH [LO | HI] [x]<cr>

where
LO = Adjustment point at the dry end (low RH).
HI = Adjustment point at the wet end (high RH). The difference

between the two points should be at least 30 %RH.

x = The reference humidity condition (%RH) that the transmitter

should be showing.

The 2-point correction is not applied immediately – you must use the

CRH SAVE command to store your adjustment to the HTM10 module.

If you have entered user adjustments using the CRH command but do not

wish to commit them, use the CRH CANCEL command.

CRH [SAVE | CANCEL]<cr>

Example (two point adjustment, low point 11 %RH and high point 75

%RH):

>pass 9000

>crh lo 11

OK

>crh hi 75

OK

>crh save

OK