Chapter 6 ____________________________________________________________ Troubleshooting

VAISALA ________________________________________________________________________ 75

CHAPTER 6

TROUBLESHOOTING

This chapter describes common problems, their probable causes and

remedies, and provides contact information for technical support.

Problem Situations

Table 11

Troubleshooting Table

Problem

Possible Causes and Solutions

Temperature reading shown by

the transmitter is too high.

1. The transmitter may be installed in an

unsuitable location, for example, near a

heat source or in sunlight. See section

Relay Configuration in DIP Mode on

page 28.

2. Check that the transmitter is installed in

proper orientation, with the arrow on the

mounting base pointing up.

Relay does not seem to be

working as configured.

1. Check DIP switch settings. Is the relay

configured using DIP switches and

rotary switch, or software? Check that

the rotary switch is not between two

positions.

2. Note the effect of hysteresis in DIP

mode; see section Relay Configuration

in DIP Mode on page 23.

3. Connect to the service port using a PC

and use the

STATUS command to view

the current relay settings.

4. Use the

RTEST command to test that

the relay is working properly.

Transmitter does not recognize

a valid serial command,

responds with message

FAIL 1: Unknown command

1. The command may be one of the

advanced commands, and you have not

enabled them using the

PASS 9000

command.

2. If you are using remote echo on the

transmitter, disable it with the

ECHO OFF command to avoid

collisions.

3. There may be an intermittent connection

problem between the transmitter and

your terminal. Issue the command again.