Reverting to factory settings using service port, Figure 33, Dip switches in factory reset position (hmw95) – Vaisala HMW90 User Manual

Page 82

Advertising
background image

User's Guide _______________________________________________________________________

80 ___________________________________________________________________ M211399EN-F

1209-028

Figure 33

DIP Switches in Factory Reset Position (HMW95)

4.

Reconnect the transmitter cover to the mounting base so it powers

up. Check the screen after power-up: when the DIP switches are in

factory reset position, you will see a notification text.

5.

Disconnect the transmitter cover again.

6.

Set the DIP switches to the positions they were before.

7.

Reconnect the transmitter cover to the mounting base. Check the

startup screens to verify the configuration.

Reverting to Factory Settings Using

Service Port

Use the FRESTORE command to restore the transmitter to factory

settings.

FRESTORE<cr>

NOTE

After using the FRESTORE command, reset the transmitter using the

RESET command.

Example:

>pass 9000

>frestore

Restoring HTM10 factory parameters

HTM10 factory parameters restored

Restoring HMW93 factory settings

72/72 parameters restored

OK

>reset

Resetting

HMW93 / 1.00.00.0000 / XM90

>

HMW95

1

ON

2 3 4 5 6 7 8

Modbus

BACnet

Parity Even

A B C

Parity None

Baud

Rate

Metric

Non-Metric

Advertising