Technical data, Specifications, Chapter 7 – Vaisala HMW90 User Manual

Page 85: Chapter 7, technical data, Ee chapter 7, Technical data, on pa

Advertising
background image

Chapter 7 _____________________________________________________________ Technical Data

VAISALA ________________________________________________________________________ 83

CHAPTER 7

TECHNICAL DATA

This chapter provides the technical data of the HMW90 series

transmitters.

Specifications

Table 14

Performance

Property

Description / Value

Relative humidity

Measurement range

0 ... 100 %RH, non-condensing

Accuracy

Temperature range +10 ... +40 °C

(+50 ... +104 °F)

0 ... 90 %RH

±1.7 %RH

90 ... 100 %RH

±2.5 %RH

Temperature range -5 ... +10 °C,

+40 ... + 55 °C (+23 ... +50 °F,

+104 ... +131°F)

0 ... 90 %RH

±3 %RH

90 ... 100 %RH

±4 %RH

Stability in typical HVAC applications

±0.5 %RH/year

Humidity sensor

Vaisala HUMICAP

®

180R

Temperature

Measurement range

-5 ... +55 °C (+23 ... +131 ºF)

Accuracy

+20 ... +30 °C (+68 ... +86 °F)

±0.2 °C (± 0.36 °F)

+10 ... +20 °C, +30 ... +40°C

(+50 ... +68 °F, +86 ... +104 °F)

±0.3 °C (± 0.54 °F)

-5 ... +10 °C, +40...+55°C

(+23 ... +50 °F, +104 ... +131 °F)

±0.5 °C (± 0.90 °F)

Temperature sensor

Digital temperature sensor

Table 15

Operating Environment

Property

Description / Value

Operating temperature range

-5 ... +55 °C (+23 ... +131 °F)

Storage temperature range

-30 ... +60 °C (-22 ... +140 °F)

Electromagnetic compliance

EN61326-1, Industrial Environment