Clear user adjustment of rh measurement, Adjust temperature measurement, Show current t adjustment – Vaisala HMW90 User Manual

Page 62: Point adjustment of t measurement

Advertising
background image

User's Guide _______________________________________________________________________

60 ___________________________________________________________________ M211399EN-F

Clear User Adjustment of RH Measurement

CRH [RESET]<cr>

Example:

>pass 9000

>crh reset

OK

Adjust Temperature Measurement

Use the CT command to perform an adjustment of the temperature

measurement. You can do a 1-point adjustment or clear the adjustment

information from the HTM10 module. Note that the factory adjustment

will remain intact when user adjustment is cleared.

Show Current T Adjustment

CT<cr>

Example (showing default temperature offset):

>pass 9000

>ct

Temperature offset : 0.000

1-point Adjustment of T Measurement

Place the transmitter in the reference condition and allow it to stabilize

before entering the adjustment.

CT [x]<cr>

where
x = The reference temperature (in degrees Celsius) that the

transmitter should be showing.

Example:

>pass 9000

>ct 23

OK