Calibration and Adjustment

HMW90 series transmitters are fully calibrated as shipped from factory.

Calibration and adjustment services are available through Vaisala Service

Centers. For contact information, see

www.vaisala.com/servicecenters

.

HMW90 series transmitters have a display that makes it easy to compare

the measured readings against any portable calibration reference. Note

that depending on the ordered configuration, the display may be hidden

under the sliding cover.

For adjustment of the measurement, you have the following options:

- 1-point adjustment using the trimmers under the sliding cover. See

section Adjustment Using Display and Trimmers on page 69.

- 1-point or 2-point adjustment using the service port. See the following

sections:

- Adjustment Using an HM70 on page 70.
- Adjustment Using a PC on page 71.

- Replacement of the Humidity and Temperature Module HTM10,

which can be ordered as a spare part. See section Replacing the

Measurement Module on page 72.

The adjustment of temperature measurement is always a simple 1-point

offset correction.

The 1-point humidity adjustment of the HMW90 series adjusts both

offset and gain, depending on the adjustment condition. In a dry

condition (for example, 11 %RH), offset is adjusted more than gain.

NOTE

The 1-point humidity adjustment requires that the target condition is at

least 50% of the currently measured value. This prevents the user from

making excessive corrections that are very unlikely to be needed.

However, if you feel that you may have done an incorrect adjustment,

you can easily remove the adjustment using the HM70 or the

CRH RESET command on the serial line.

NOTE

Only T adjustment is available on the TMW92 and TMW93 transmitters.