76 ___________________________________________________________________ M211399EN-F

Error Messages

Error Messages on the Display

Table 12

Error Messages on the Display

Error Message

Possible Cause and Solution

HTM10 01

HTM10 02

Communication failure with HTM10

module. Reconnect the module and check that it sits

firmly in place.

HTM10 03

HTM10 04

HTM10 05

HTM10 06

Problem with the HTM10 module.

1. Check the module for damage.

2. Check if the humidity sensor is missing or loose. Note

that the humidity sensor is not included on TMW-type

transmitters since they only measure temperature.

3. Replace the module if unable to remove the problem.

Internal 1

Internal 2

Internal 3

Internal problem with the transmitter.

1. Reset the transmitter.

2. Restore the factory settings using service port or DIP

switches if reset does not help.

Error Messages on the Serial Line

View Currently Active Errors

Use the ERRS command to view currently active errors on the serial

line:

ERRS<cr>

Example:

>errs

NO ERRORS

Advertising