AM Gain Calibration

4-40

AM Gain Calibration

Description

This calibration adjusts the gain of the AM circuitry to provide a 10 dB drop for a 1 V input signal. First a
power level is set and a one-volt signal is connected to the input. AM is enabled and the AM DACs are
adjusted for an exact 10 dB drop. The resulting DAC value is stored as a calibration array.

Required Test Equipment

• Agilent E4418B E-Series power meter

• Agilent E9304A E-Series power sensor

• Agilent 33120A function generator

• Agilent 3458A Option 002 digital multimeter

Procedure

Figure 4-31

AM Gain Calibration Setup

1. Connect the equipment as shown above.

2. Preset all of the equipment.

3. Follow the instructions as they appear on the controller’s display.