Performance Tests

Internal FM Accuracy and Distortion

3- 9

Internal FM Accuracy and Distortion

These automated tests verify the FM accuracy and distortion specifications. Accuracy is verified directly with
a measuring receiver, and distortion is verified by measuring the demodulated output of the measuring
receiver, using the audio analyzer. The audio analyzer provides a more accurate verification with better
resolution, as opposed to the measuring receiver alone.

Recommended Equipment

• Agilent 8902A measuring receiver

• Agilent 8903B audio analyzer with Options 010 and 051, or Options 011 and 050

Equipment Setup

Figure 3-6

Internal FM Accuracy and Distortion Setup