Performance Tests

AM Frequency Response

3-14

AM Frequency Response

This automated test verifies the AM frequency response specifications. The equipment measures the
variations in modulation depth due to changes in the applied AM rate; dc to 10 kHz. The response is relative
to a 1 kHz rate set at the test depth, and expressed in dB. Each frequency is tested with the following
sequence:

1. A reference signal is measured.

2. The set depth is measured for each applied rate change.

3. Each depth measurement is compared to the reference signal depth.

The comparison is a calculation of the difference in the depths relative to the reference rate. The relative
value is the AM Frequency Response.

Recommended Equipment

• Agilent 8902A measuring receiver

• Agilent 8904A function generator

• Use one of the following:

— for models with Option 501, 502, 503, 504 or Option UNJ, use an Agilent 8662A/8663A

high-performance RF signal generator

— for models with Option 506, use an Agilent E4426/E4437 signal generator

• MD/MDC-164 mixer (0.5 to 9 GHz)

Equipment Setups

Figure 3-12

AM Frequency Response

1300 MHz Setup

Advertising