Performance Tests

Overview of Performance Tests

3-4

Running Performance Tests

1. Load the Service Support Software into the personal computer. Refer to

Chapter 1, “Getting Started,”

for

installation instructions.

2. Click on the installed Service Support Software icon.

3. Fill in the fields in the User Information window (

Figure 3-1

):

a. In the User Name field (item 1), enter User. (Case sensitive.)

b. In the Password field (item 2), enter User. (Case sensitive.)

4. Click OK.

Figure 3-1

The User Information Window

5. In the DUT (Device Under Test) Selection Window (

Figure 3-2

):

a. In the Select Model list (item 3), choose the correct model number of your signal generator.

b. In the Serial Number box (item 4), enter the complete serial number of the DUT.

c. In the Address box (item 5), enter the two-digit GPIB address of the signal generator. (To display the

address on the signal generator, press

Utility

>

GPIB/RS-232

>

GPIB Address.

The

GPIB Address

softkey

shows the current address.)

6. Click OK.