Performance Tests

Dual Arbitrary Waveform Generator Check

3-40

out of phase and that they do not have any discontinuities. Refer to the illustration below.

• Frequency is approximately 64 kHz.

• Amplitude is approximately 1.9 V

p-p

.

• Both waveforms are symmetrical around 0 volts (the average voltage is approximately 0V).