Power level accuracy, Recommended equipment, Power level accuracy -21 – Agilent Technologies E4438C User Manual

Page 69: Recommended equipment -21

Advertising
background image

Performance Tests

Power Level Accuracy

3- 21

Power Level Accuracy

A power meter is used to verify performance over the +13 to

45 dBm range. The absolute power level

measured at

40 dBm is used as a reference for all lower level measurements.

For power levels below

45 dBm, a vector signal analyzer is used to make relative power measurements. A

low noise amplifier (LNA) and step attenuator are connected in series to control the absolute power level
input to the signal analyzer. For power level settings

≥ −

90 dBm, the step attenuator is set to approximately

cancel the gain of the LNA. Below

90 dBm, the step attenuator is set to 0 dB, which keeps the signal level

well above the signal analyzer’s noise floor.

Relative measurements are accomplished by setting the DUT amplitude. For example:

Set the DUT to a specific frequency, and an amplitude of

40 dBm. For the purpose of this example, assume

the power meter measures

40.1 dBm at these settings. The signal analyzer does a peak search, sets the

reference level, and sets the marker to delta mode (0 dB). The DUT amplitude then changes to

50 dBm, and

the signal analyzer makes a relative power measurement (assume it to be

10.15 dB). The absolute power

measured is

40.1 dBm + (

10.15 dB), or

50.25 dBm.

To determine power level accuracy below

90 dBm, the signal analyzer makes the 0 dB marker delta

measurement with the amplitude set to

90 dBm, and all lower level measurements are made relative to this

reference. The previously measured error for the

90 dBm setting is then accounted for in the measurement.

For frequencies above 2 GHz, an Agilent 8563EC Option 001 spectrum analyzer is used to downconvert the
frequency to 321.4 MHz.

Recommended Equipment

• Agilent E4418B E-Series power meter

• Agilent E9304A E-Series power sensor

• Agilent 89441A vector signal analyzer with Options AYA, AYH, AY7, AY9, B7A, UFG or UTH

• Agilent 8563EC Option 001 spectrum analyzer

• Agilent 8495G programmable step attenuator (0 to 70 dB)

• Agilent 11713A step attenuator driver

See

Chapter 2, “Required Equipment,” on page 2-1

for details on the following:

— Low Frequency Low Noise Amplifier (LNA)

— High Frequency Low Noise Amplifier (LNA)

Advertising