Agilent Technologies 01664-97005 User Manual

Page 119

Advertising
background image

7

Using four 6-by-2 test connectors, four BNC Couplers, and four SMA (m) - BNC (f)
Adapters, connect the logic analyzer to the pulse generator channel outputs. To
make the test connectors, see chapter 3, "Testing Performance."

a

Connect the even-numbered channels of the lower byte of the pod under test to the
pulse generator channel 1 Output and J-clock.

b

Connect the odd-numbered channels of the lower byte of the pod under test to the
pulse generator channel 1 Output.

c

Connect the even-numbered channels of the upper byte of the pod under test and the
clock channel to the pulse generator channel 2 Output.

d

Connect the odd-numbered channels of the upper byte of the pod under test to the
pulse generator channel 2 Output.

8

On the logic analyzer, press Run. The display should look similar to the figure below.

9

If the display looks like the figure, then the cable passed the test.

If the display does not look similar to the figure, then there is a possible problem
with the cable or probe tip assembly. Causes for cable test failures include the
following:

open channel.

channel shorted to a neighboring channel.

channel shorted to either ground or a supply voltage.

Return to the troubleshooting flowchart.

Troubleshooting

To test the logic analyzer probe cables

5–31