Agilent Technologies 01664-97005 User Manual

Page 75

Advertising
background image

6

Test the next clock.

a

Press the Format key, then select Master Clock.

b

Turn off and disconnect the clock just tested.

c

Repeat steps 3 and 5 for the next clock listed in the table in step 3, until all listed clock
edges have been tested.

7

Test the next setup/hold combination. a Press the Format key, then select Master
Clock and turn off the clock just tested. b Repeat steps 1 through 7 for the next
setup/hold combination listed in step 1, until all listed setup/hold combinations have
been tested.

When aligning the data and clock waveforms using the oscilloscope, align the waveforms
according to the setup time of the setup/hold combination being tested, +0.0 ps or -100.0 ps.

Testing Performance

To test the single-clock, multiple-edge, state acquisition

3–53