To test the threshold accuracy, Set up the equipment – Agilent Technologies 01664-97005 User Manual

Page 30

Advertising
background image

To test the threshold accuracy

Testing the threshold accuracy verifies the performance of the following specification:

Clock and data channel threshold accuracy.

These instructions include detailed steps for testing the threshold settings of pod 1.
After testing pod 1, connect and test pod 2. To test pod 2, follow the detailed steps
for pod 1, substituting the pod 2 for pod 1 in the instructions.

Each threshold test tells you to record the voltage reading in the performance test
record located at the end of this chapter. To check if each test passed, verify that the
voltage reading you record is within the limits listed on the performance test record.

Equipment Required

Equipment

Critical Specifications

Recommended Model/Part

Digital Multimeter

0.1 mV resolution, 0.005% accuracy

Agilent 3458A

Function Generator

DC offset voltage

±6.3 V

Agilent 3325B Option 002

BNC-Banana Cable

Agilent 11001-60001

BNC Tee

Agilent 1250-0781

BNC Cable

Agilent 8120-1840

BNC Test Connector, 17x2

Set up the equipment

1

Turn on the equipment required and the logic analyzer. Let them warm up for
30 minutes before beginning the test.

2

Set up the function generator.

a

Set up the function generator to provide a DC offset voltage at the Main Signal output.

b

Disable any AC voltage to the function generator output, and enable the high voltage
output.

c

Monitor the function generator DC output voltage with the multimeter.

3–8

Advertising