To test the multiple-clock, multiple-edge, state acquisition

Testing the multiple-clock, multiple-edge, state acquisition verifies the performance
of the following specifications:

Minimum master to master clock time.

Maximum state acquisition speed.

Setup/Hold time for multiple-clock, multiple-edge, state acquisition.

Minimum clock pulse width.

This test checks data using multiple clocks at three selected setup/hold times.

Equipment Required

Equipment

Critical Specifications

Recommended
Model/Part

Pulse Generator

100 MHz 3.5 ns pulse width, < 600 ps rise time

Agilent 8131A option 020

Digitizing Oscilloscope

≥ 6 GHz bandwidth, < 58 ps rise time

Agilent 54121T

Adapter

SMA(m)-BNC(f)

Agilent 1250-1200

SMA Coax Cable (Qty 3)

18 GHz bandwidth

Agilent 8120-4948

BNC Cable

BNC(m)(m) 48 in. >2 GHz bandwidth

Agilent 8120-1840

Coupler

BNC(m)(m)

Agilent 1250-0216

BNC Test Connector,
6x2 (Qty 4)

Set up the equipment

1

Turn on the equipment required and the logic analyzer. Let them warm up for
30 minutes before beginning the test if you have not already done so.

2

Set up the pulse generator.

a

Set up the pulse generator according to the following table.

Pulse Generator Setup

Channel 1

Channel 2

Period

Delay: 0 ps

Doub: 20.0 ns

40 ns

Width: 4.5 ns

Width: 3.5 ns

High:

−0.9 V

High:

−0.9 V

Low:

−1.7 V

Low:

−1.7 V

b

Disable the pulse generator channel 2 COMP (with the LED off).

3–34