Adjust the pulse generator channel 1 Delay, then select Precision Edge Find in the
oscilloscope Delta T menu. Repeat this step until the pulses are aligned according to
the setup time of the setup/hold combination selected, +0.0 ps or

−100 ps.

4

Select the clocks to be tested.

a Select the clock field to be tested and then select J

↑ + K↑ as the clock edges.

b

Select Done to exit the Master Clock menu.

Testing Performance

To test the multiple-clock, multiple-edge, state acquisition

3–41