Performance Test Record

Performance Test Record

1664A Logic Analyzer_______

Serial No.______________________

Work Order No.___________________

Recommended Test Interval - 2 Year/4000 hours

Date___________________

Recommended next testing___________________

Temperature___________________

Test

Settings

Results

Self-Tests

Pass/Fail

________

Threshold
Accuracy

± (100 mV + 3% of
threshold setting)

Limits

Measured

Pod 1

TTL

, ±145 mV

ECL,

±139 mV

-User,

±280 mV

+User,

±280 mV

0 V,

±100 mV

TTL VL
TTL VH
ECL VL
ECL VH
-User VL
- User VH
+ User VL
+ User VH
0 V User VL
0 V User VH

+1.355 V
+1.645 V
-1.439 V
-1.161 V
-6.280 V
-5.720 V
+5.720 V
+6.280 V
-100 mV
+100 mV

________
________
________
________
________
________
________
________
________
________

Pod 2

TTL

, ±145 mV

ECL,

±139 mV

-User,

±280 mV

+User,

±280 mV

0 V,

±100 mV

TTL VL
TTL VH
ECL VL
ECL VH
-User VL
- User VH
+ User VL
+ User VH
0 V User VL
0 V User VH

+1.355 V
+1.645 V
-1.439 V
-1.161 V
-6.280 V
-5.720 V
+5.720 V
+6.280 V
-100 mV
+100 mV

________
________
________
________
________
________
________
________
________
________

Glitch Capture

Minimum Detectable
Glitch 3.5 ns

Pass/Fail

Pod 1

________

Pod 2

________

3–59