Agilent Technologies 01664-97005 User Manual

Page 62

Advertising
background image

Check the setup/hold with single clock edges, multiple clocks

1

Select the logic analyzer setup/hold time.

a

In the logic analyzer Format menu, select Master Clock.

b

Select and activate any two clock edges.

c

Select the Setup/Hold field and select the setup/hold to be tested for all pods. The first
time through this test, use the top combination in the following table.

Setup/Hold Combinations

4.5/0.0 ns

0.0/4.5 ns

2.0/2.5 ns

d

Select Done to exit the setup/hold combinations.

2

Disable the pulse generator channel 2 COMP (with the LED off).

3

Using the Delay mode of the pulse generator channel 1, position the pulses
according to the setup time of the setup/hold combination selected, +0.0 ps or

−100

ps.

a

In the oscilloscope Delta V menu, set the Marker 1 Position to Chan 1, then set
Marker 1 at

−1.3000 V. Set the Marker 2 Position to Chan 2, then set Marker 2 at

−1.3000 V.

b

In the oscilloscope Delta T menu, select Start on Pos Edge 1. Select Stop on Pos
Edge 1.

Testing Performance
To test the multiple-clock, multiple-edge, state acquisition

3–40