Connect the logic analyzer – Agilent Technologies 01664-97005 User Manual

Page 48

Advertising
background image

Connect the logic analyzer

1

Using the 6-by-2 test connectors, connect the logic analyzer clock and data channels
listed in the following table to the pulse generator. Install a BNC cable between the
pulse generator channel 2 output and the 6x2 test connector with the logic analyzer
clock leads.

2

Using SMA cables, connect the oscilloscope to the pulse generator channel 1
Output, channel 2 Output, and Trig Output.

Connect the 1664A Logic Analyzer to the Pulse Generator

Testing
Combination

Connect to
Agilent 8131A
Channel 1 Output

Connect to
Agilent 8131A
Channel 1 Output

Connect to
Agilent 8131A
Channel 2 Output

1

Pod 1, channel 3
Pod 2, channel 3

Pod 1, channel 11
Pod 2, channel 11

J-clock

3

Activate the data channels that are connected according to the previous table.

a

Press the Format key.

b

Select the field showing the channel assignments for one of the pods being tested,
then press the Clear entry key. Using the arrow keys, move the selector to the data
channels to be tested, then press the Select key. An asterisk means that a channel is
turned on. When all the correct channels of the pod are turned on, press the Done
key. Follow this step for the remaining pods.

Testing Performance
To test the single-clock, single-edge, state acquisition

3–26