To test the time interval accuracy

Testing the time interval accuracy does not check a specification, but does check the
following:

125 MHz oscillator

This test verifies that the 125 MHz timing acquisition synchronizing oscillator is
operating within limits.

Equipment Required

Equipment

Critical Specifications

Recommended
Model/Part

Pulse Generator

100 MHz 3.5 ns pulse width, < 600 ps rise time

Agilent 8131A Option 020

Function Generator

Accuracy

≤ (5)(10

−6

)

× frequency

Agilent 3325B Option 002

SMA Cable

Agilent 8120-4948

Adapter

BNC(m)-SMA(f)

Agilent 1250-2015

BNC Test Connector,
6x2

Set up the equipment

1

Turn on the equipment required and the logic analyzer. Let them warm up for
30 minutes if you have not already done so.

2

Set up the pulse generator according to the following table.

Pulse Generator Setup

Channel 1

Period

Mode

EXT TRIG

Delay: 0 ps

5 us

TRIG

Slope: Positive

Width: 2.5 us

THRE: 1.0 V

High:

−0.9 V

Low:

−1.7 V

COMP: Disabled
(LED off)