5

Note: This step is only done the first time through the test, to create a Compare file.
For subsequent runs, go to step 6. Use the following to create a Compare file:

a

Press Run. The display should show a checkerboard pattern of alternating As and 5s.
Verify the pattern by scrolling through the display.

b

Press the List key. In the pop up menu, use the RPG knob to move the cursor to
Compare. Press Select.

c

In the Compare menu, move the cursor to Copy Listing to Reference, then select
Execute in the pop-up menu and press the Select key.

d

Move the cursor to Specify Stop Measurement and press the Select key. Press Select
again to turn on Compare. At the pop up menu, select Compare. Move the cursor to
the Equal field and press the Select key. At the pop up menu, select Not Equal. Press
Done.

e

Move the cursor to the Reference Listing field and select. The field should toggle to
Difference Listing.

6

Press the blue shift key, then press the Run key. If 2 - 4 acquisitions are obtained
without the "Stop Condition Satisfied" message appearing, then the test passes.
Press Stop to halt the acquisition. Record the Pass or Fail results in the
performance test record.

7

Test the next clock.

a

Press the Format key, then select Master Clock.

b

Turn off and disconnect the clock just tested.

c

Repeat steps 4, 6, and 7 for the next clock edge listed in the table in step 4, until all
listed clock edges have been tested.

8

Enable the pulse generator channel 2 COMP (with the LED on).

9

Check the clock pulse width.

a

Enable the pulse generator channel 1 and channel 2 outputs.

b

In the oscilloscope Timebase menu, select Delay. Using the oscilloscope knob,
position the clock waveform so that the waveform is centered on the screen.

c

In the oscilloscope Delta V menu, set the Marker 1 Position to Chan 2, then set
Marker 1 at

−1.3000 V. Set the Marker 2 Position to Chan 2, then set Marker 2 at

−1.3000 V.

d

In the oscilloscope Delta T menu, select Start On Neg Edge 1. Select Stop on Pos
Edge 1.

e

If the pulse width is outside the limits, adjust the pulse generator channel 2 width and
select the oscilloscope Precision Edge Find until the pulse width is within limits.

Testing Performance

To test the single-clock, single-edge, state acquisition

3–31