Set up the equipment – Agilent Technologies 01664-97005 User Manual

Page 45

Advertising
background image

To test the single-clock, single-edge, state acquisition

Testing the single-clock, single-edge, state acquisition verifies the performance of the
following specifications:

Minimum master to master clock time.

Maximum state acquisition speed.

Setup/Hold time for single-clock, single-edge, state acquisition.

Minimum clock pulse width.

This test checks the data channels using a single-edge clock at three selected
setup/hold times.

Equipment Required

Equipment

Critical Specifications

Recommended Model/Part

Pulse Generator

100 MHz 3.5 ns pulse width, < 600 ps rise time

Agilent 8131A option 020

Digitizing Oscilloscope

≥ 6 GHz bandwidth, < 58 ps rise time

Agilent 54121T

Adapter

SMA(m)-BNC(f)

Agilent 1250-1200

SMA Coax Cable (Qty 3)

18 GHz bandwidth

Agilent 8120-4948

BNC Cable

BNC(m)(m) 48 in. >2 GHz bandwidth

Agilent 8120-1840

Coupler

BNC(m)(m)

Agilent 1250-0216

BNC Test Connector,
6x2 (Qty 4)

Set up the equipment

1

Turn on the equipment required and the logic analyzer. Let them warm up for
30 minutes before beginning the test if you have not already done so.

2

Set up the pulse generator.

a

Set up the pulse generator according to the following table.

Pulse Generator Setup

Channel 1

Channel 2

Period

Delay: 0 ps

Doub: 20.0 ns

40 ns

Width: 3.5 ns

Width: 3.5 ns

High: -0.9 V

High: -0.9 V

Low: -1.7 V

Low: -1.7 V

b

Disable the pulse generator channel 2 COMP (with the LED off).

3–23