To test the single-clock, multiple-edge, state acquisition

Testing the single-clock, multiple-edge, state acquisition verifies the performance of
the following specifications:

Minimum master to master clock time.

Maximum state acquisition speed.

Setup/Hold time for single-clock, multiple-edge, state acquisition.

This test checks data channels using a multiple-edge single clock at three selected
setup/hold times.

Equipment Required

Equipment

Critical Specifications

Recommended
Model/Part

Pulse Generator

100 MHz 3.5 ns pulse width, < 600 ps rise time

Agilent 8131A option 020

Digitizing Oscilloscope

≥ 6 GHz bandwidth, < 58 ps rise time

Agilent 54121T

Adapter

SMA(m)-BNC(f)

Agilent 1250-1200

SMA Coax Cable (Qty 3)

18 GHz bandwidth

Agilent 8120-4948

BNC Cable

BNC(m)(m) 48 in. >2 GHz bandwidth

Agilent 8120-1840

Coupler

BNC(m)(m)

Agilent 1250-0216

BNC Test Connector,
6x2 (Qty 4)

Set up the equipment

1

Turn on the equipment required and the logic analyzer. Let them warm up for
30 minutes before beginning the test if you have not already done so.

2

Set up the pulse generator according to the following table.

Pulse Generator Setup

Channel 1

Channel 2

Period

Doub: 0 ps

Delay: 0 ps

40 ns

Width: 4.0 ns

Dcyc: 50%

High:

−0.9 V

High:

−0.9 V

Low:

−1.7 V

Low:

−1.7 V

COMP: Disabled
(LED off)

COMP: Disabled
(LED off)

3–45