Check the setup/hold combination – Agilent Technologies 01664-97005 User Manual

Page 73

Advertising
background image

Check the setup/hold combination

1

Select the logic analyzer setup/hold time.

a

In the logic analyzer Format menu, select Master Clock.

N O T E

The first time through this test, assign the clocks according to the first testing combination in
step 3 of these procedures.

b

In the Master Clock menu, select Setup/Hold.

c

In the Setup/Hold menu, select the setup/hold field, then select for all pods the
setup/hold combination to be tested. The first time through this test, use the top
combination in the following table.

Setup/Hold Combinations

4.0/0.0 ns

0.0/4.0 ns

2.0/2.0 ns

d

Select Done to exit the setup/hold combinations.

2

Using the Delay mode of the pulse generator Channel 2, position the pulses
according to the setup time of the setup/hold combination selected, +0.0 ps or
-100.0 ps.

a

In the oscilloscope Delta V menu, set the Marker 1 Position to Chan 1, then set Marker
1 at -1.3000 V. Set the Marker 2 Position to Chan 2, then set Marker 2 at -1.3000 V.

b

In the oscilloscope Delta T menu, select Start on Pos Edge 1. Select Stop on Pos
Edge 1. Select Precision Edge Find.

Testing Performance

To test the single-clock, multiple-edge, state acquisition

3–51