To test the auxiliary power

The +5 V auxiliary power is protected by a current overload protection device. If the current
on pins 1 and 39 exceed 0.33 amps, the circuit will open. When the short is removed, the
circuit will reset in approximately 1 minute. There should be +5 V after the 1 minute reset
time.

Equipment Required

Equipment

Critical Specifications

Recommended
Model/Part

Digital Multimeter

0.1 mV resolution, better than
0.005% accuracy

Agilent 3478A

Using the multimeter, verify the +5 V on pins 1 and 39 of the probe cables.

Troubleshooting
To test the auxiliary power

5–32