Agilent Technologies 01664-97005 User Manual
Page 43
Advertising
3
Set up the Trigger menu.
a
Press the Trigger key.
b
Select Modify Trigger, then Clear Trigger, then select All in the pop-up menus.
4
Using the Precision Edge Find in the Delta T menu of the oscilloscope, verify that
the pulse widths of the pulse generator channels 1 and 2 are 3.450 ns, +50 ps or
−100
ps. If necessary, adjust the pulse widths of the pulse generator channels 1 and 2.
5
Set up the Waveform menu to view all the channels.
a
Select one of the Glitch labels, then select Delete All in the pop-up menu.
b
Select All, then select continue.
c
Press the Select key, then select Insert in the pop-up menu.
d
Press the Select key, then select Sequential in the pop-up menu.
Testing Performance
To test the glitch capture
3–21