Agilent Technologies 01664-97005 User Manual

Page 43

Advertising
background image

3

Set up the Trigger menu.

a

Press the Trigger key.

b

Select Modify Trigger, then Clear Trigger, then select All in the pop-up menus.

4

Using the Precision Edge Find in the Delta T menu of the oscilloscope, verify that
the pulse widths of the pulse generator channels 1 and 2 are 3.450 ns, +50 ps or

−100

ps. If necessary, adjust the pulse widths of the pulse generator channels 1 and 2.

5

Set up the Waveform menu to view all the channels.

a

Select one of the Glitch labels, then select Delete All in the pop-up menu.

b

Select All, then select continue.

c

Press the Select key, then select Insert in the pop-up menu.

d

Press the Select key, then select Sequential in the pop-up menu.

Testing Performance

To test the glitch capture

3–21