Test the − user threshold – Agilent Technologies 01664-97005 User Manual

Page 35

Advertising
background image

Test the

− User threshold

1

Move the cursor to the field to the right of Pod 1. Type –6.00, then use the left and
right cursor control keys to highlight V. Press the Select key.

2

On the function generator front panel, enter

−5.718 V ±1 mV DC offset. Use the

multimeter to verify the voltage.

The activity indicators for pod 1 should show all data channels and the J-clock channel at a
logic high.

3

Using the Modify down arrow on the function generator, decrease offset voltage in
1-mV increments until all activity indicators for pod 1 show the channels at a logic
low. Record the function generator voltage in the performance test record.

4

Using the Modify up arrow on the function generator, increase offset voltage in
1-mV increments until all activity indicators show the channels at a logic high.
Record the function generator voltage in the performance test record.

Testing Performance

To test the threshold accuracy

3–13

Advertising