Set up the logic analyzer, Connect the logic analyzer – Agilent Technologies 01664-97005 User Manual

Page 40

Advertising
background image

3

Set up the oscilloscope.

Oscilloscope Setup

Time Base

Display

Delta V

Delta T

Time/Div: 1.00 ns/div

mode: avg

V markers on

T markers on

delay: 17.7000 ns

# of avg: 16

marker 1 position: Chan 1

start on: Pos Edge 1

screens: dual

marker 2 position: Chan 2

stop on: Pos Edge 1

Channel

Channel 1

Channel 2

Display

on

on

Probe Atten

20.00

20.00

Volts/Div

400 mV

400 mV

Offset

−1.3000 V

−1.3000 V

Set up the logic analyzer

1

Press the Config key. Assign all pod fields to Machine 1. To assign the pod fields,
select the pod fields, then select Machine 1 in the pop-up menu.

2

In the Analyzer 1 box, select the Type field. Select Timing in the pop-up menu.

Connect the logic analyzer

1

Using SMA cables, connect the oscilloscope to the pulse generator channel 1
Output, channel 2 Output, and Trig Output.

2

Using the 6-by-2 test connectors, connect the first combination of logic analyzer
clock and data channels listed in the table to the pulse generator.

You will repeat this test for the remaining combinations.

Testing Performance
To test the glitch capture

3–18