To test the glitch capture, Set up the equipment – Agilent Technologies 01664-97005 User Manual

Page 39

Advertising
background image

To test the glitch capture

Testing the glitch capture verifies the performance of the following specification:

Minimum detectable glitch.

This test checks the minimum detectable glitch on sixteen data channels at a time.

Equipment Required

Equipment

Critical Specifications

Recommended Model/Part

Pulse Generator

100 MHz 3.5 ns pulse width, < 600 ps rise time

Agilent 8131A Option 020

Digitizing Oscilloscope

≥ 6 GHz bandwidth , < 58 ps rise time

Agilent 54121T

SMA Coax
(Qty 3)

18 GHz bandwidth

Agilent 8120-4948

Adapter (Qty 4)

SMA(m)-BNC(f)

Agilent 1250-1200

Coupler (Qty 4)

BNC (m)(m)

Agilent 1250-0216

BNC Test Connector, 6x2 (Qty 4)

Set up the equipment

1

Turn on the equipment required and the logic analyzer. Let them warm up for
30 minutes before beginning the test if you have not already done so.

2

Set up the pulse generator.

Pulse Generator Setup

Channel 1

Channel 2

Period

Delay: 0 ps

Delay : 0 ps

22.0 ns

Width: 3.5 ns

Width: 3.5 ns

High:

−0.9 V

High:

−0.9 V

Low:

−1.7 V

Low:

−1.7 V

3–17