6010 Series Operations Manual

Rev E 2/14/2012

Page 19 of 68

Connecting Communications or Analog Outputs (DLUX301 & DLUX302)

With the purchase of DLUX301 or DLUX302, the controllers are upgraded with

additional features such as RS485 communications, and analog outputs. A set of

terminals are provided to connect to RS485 communications and analog outputs.

Analog Outputs

DLUX301 & DLUX302 provide analog outputs as either milliamps (0-20mA, 4-20mA) or

voltage (0-5V, 1-5V, 0-10V, 0-20V) signal output that represents each of the displayed

temperature (and humidity) values. These options can be used for connection to in-

house data acquisition, recorder, or alarm system.

Factory default settings are as follows:

Parameter

Analog Output

Corresponding Value

Temperature

0 – 5 V

0 – 100 °C

Humidity

0 – 5 V

0 – 100 %rh

Connect shielded wires to the appropriate signal terminals: I(+) for current (mA) or V(+)

for voltage (DC). For both current and voltage outputs, COM(-) is common terminal.

The controller itself must be programmed for the voltage or current signal and

corresponding scale (see Operations Section)

Communications

RS-485

Communications

Connections

Analog Output

Connections