The temperature and humidity controllers are factory

programmed for precise control. Unlocking the

controllers gives the user access to all parameters.

Modifying parameters that are not thoroughly

understood can adversely affect chamber performance

that will not be covered under warranty.

SECTION 14 – ADVANCED USERS SECTION

Unlocking the Controllers

To unlock an individual controller

1) Press and hold the advance and infinity keys simultaneously for six seconds

2) Press the up key until LOC is displayed in the upper display

3) Press the advance key until rLoC is displayed in the lower display

4) Press the up key to change the security level from 2 to 5

5) Press the advance key until sLoC is displayed in the lower display

6) Press the up key to change the security level from 2 to 5

7) Press infinity key twice to return to the main menu

All controller parameters are now available to be modified. Once the appropriate

changes have been made, it is highly recommended to relock the controllers per the

instructions below.

Locking the Controllers

To lock an individual controller

1) Press and hold the advance and infinity keys simultaneously for six seconds

2) Press the up key until LOC is displayed in the upper display

3) Press the advance key until rLoC is displayed in the lower display

4) Press the up key to change the security level from 5 to 2

5) Press the advance key until sLoC is displayed in the lower display

6) Press the up key to change the security level from 5 to 2

7) Press infinity key twice to return to main menu

All controller parameters are now locked.