SECTION 5 – OPERATION

With the chamber properly installed and the appropriate utilities connected, the power

switch on the upper left of the control bezel can be turned on. Within a few minutes, the

temperature and humidity will begin to approach set-points. Allow the unit to stabilize for

12 hours before use or prior to making any calibration adjustments.

Changing the Temperature Set-point

The standard cabinet has a temperature set-point range of 5C to 70C. With the

purchase of and optional defrost package (DEFR201), the low end temperature range

can be extended to 0C.

To set the temperature set-point, press the UP arrow to increase the temperature set-

point by 0.1

C. Press the DOWN arrow to decrease the temperature set-point by 0.1

C.

Pressing and holding either button will cause the set-point to scroll rapidly in either

direction.

To turn off the temperature control system, but still display the chamber temperature

level, press the EZ button. The words “OFF” will display in the set-point area. To toggle

back to controlled temperature, press the EZ button again.

If the unit is equipped with the deluxe controller package (DLUX301 or DLUX302),

additional ramp and soak features are available. Refer to Section 7 of the manual to

program these features.

Temperature

Set-point

Actual

Temperature

Increase

Temperature

Set-point

Decrease

Temperature

Set-point

EZ Button