6010 Series Operations Manual

Rev E 2/14/2012

Page 60 of 68

SECTION 12 – TROUBLESHOOTING

Problem -- Unit will not turn on

Is the unit connected to a dedicated electrical circuit as defined in the installation

section of the manual?

Is there power at the electric outlet the unit is plugged into?

Is the unit’s power switch turned on?

Problem -- Unit temperature is above / below temperature set-point

Has the unit’s temperature set-point been recently lowered / raised and if so has the

unit been allowed 12 hours stabilize at the new set-point?

Has the inner door been recently opened for an extended period of time?

Is the access port stopper in the right side of the cabinet installed?

Is the condenser filter on the front of the cabinet clean?

Unit humidity level is above / below humidity set-point

Is the unit connected to a water source as specified in the installation section of the

manual?

Has the unit been leveled to insure the cabinet drain works correctly?

The cabinet’s drain line uses gravity to remove water. Does the drain line have any

rises in it above the cabinet’s drain level that could be trapping water?

Has the unit’s humidity set-point been recently lowered / raised and if so has the unit

been allowed time to stabilize at the new set-point?

Has the inner door been recently opened for an extended period of time?

Is the access port stopper in the right side of the cabinet installed?

Is the condenser filter on the front of the cabinet clean?