Changing the Humidity Set-point

The standard cabinet has a humidity set-point range of 20% to 98% relative humidity.

With the purchase of an optional dryer package (DRYR301), the low end humidity range

can be extended. For information on the operation of DRYR301, refer to the

accessories section.

To set the humidity set-point, press the UP arrow to increase the humidity set-point by

1% RH. Press the DOWN arrow to decrease the humidity set-point by 1% RH.

Pressing and holding either button will cause the set-point to scroll rapidly in either

direction.

If humidity is not required in the chamber, the RH system can be disabled by turning off

the Enable Humidity switch. This will disable the entire RH system including the

controller. The system can be turned back on at anytime.

To turn off the humidity control system, but still display the chamber humidity level, press

the EZ button. The words “OFF” will display in the set-point area. To toggle back to

controlled humidity, press the EZ button again.