Installing the Heatless Dryer Package (DRYR301)

The Heatless Dryer Package extends the operational limits to a minimum humidity

control point of 2% RH and improves the dehumidification rate by purging the chamber

with dry air. Air flows through the Dryer Package only when dehumidification is required

to maintain the humidity set point.

It uses compressed air at 90 to 100 psig to operate. Maximum airflow required is 480

SCFH (8 CFM). Of this 300 SCFH (5 CFM) is injected into the chamber as dry air and

180 SCFH (3 CFM) is exhausted into the room after being used to regenerate the

desiccant. Particles of dirt or rust, as well as large amounts of condensed moisture or oil

that may be in the compressed air line feeding the dryer must be removed from the air

stream before it enters the dryer.

Air Supply

Inlet

Tower Clamps

(optional)

Power Cord

Air Tube

Coalescent Filter

With Drain

Pressure

Regulator Dial

Pressure

Regulator Knob

Flow Meter

Adjustment

Mounting

Brackets

Flow Meter

Outlet