Caron 6017 User Manual

Page 65

Advertising
background image

6010 Series Operations Manual

Rev E 2/14/2012

Page 65 of 68

8. Select the step type by pressing the ADVANCE button.

9. Use the UP and DOWN arrow buttons to define each parameter and the

ADVANCE button to go to the next parameter. The parameters are described

below:

Parameter

Description

Display

Target set-point Set point for this step

t9.SP

Hours

Number of hours for a timed step

hoUr

Minutes

Number of minutes for a timed step

Min

Seconds

Number of seconds for a timed step

SEC

Rate

Ramping rate in ºC or %rh per minute

rAtE

Wait for process

instance

Instance for wait for process step

W.Pi

Wait for process

value

Set-point to wait until reached

WPr

Wait Event 1

Event to wait until reached

WE.1

Wait Event 2

Event to wait until reached

WE.2

Jump Step

Step to jump to

JS

Jump Count

Number of jumps; a 0 indicates an infinite number

JC

End Type

What controller will do when profile ends

End

Off

turn control outputs off when profile ends

oFF

Hold

hold the last set-point when profile ends

HoLd

User

revert to previous set-point when profile ends USEr

Event Output 1

Whether event 1 is on or off during this step

Ent1

Event Output 2

Whether event 2 is on or off during this step

Ent2

10.

Repeat steps 3 through 9 until the entire profile has been programmed

11.

Press and hold the INFINITY button for 2 seconds to exit to home page

12.

“Lock” the controller (if desired) by referring to Section 14

For the previous profile example, here is how the entire 7 step program would look like

Parameter

Parameter

Display

Variable

Display

Profile 1

FiLE

P1

Step 1

P1

1

Timed step type

S.tyP

ti

Target set-point (20ºC)

T9.SP

20

Time in hours

hoUr

0

Time in minutes

Min

0

Time in seconds

SEC

0

Advertising