6010 Series Operations Manual

Rev E 2/14/2012

Page 48 of 68

move off of the chart allowing you to place the recording chart paper onto the recorder.

Press and release the Change Chart button to begin recording.

Changing the Chart Paper:

Press and hold the Change Chart button (#3) for approximately one (1) second until the

pen begins to move off scale and then release the button.

Note: The green LED light will flash fast while the thermal pen arm is moving off scale.

Wait until the thermal pen arm has moved completely off scale and stops (the green

LED light will stop flashing and will be steady On). To remove the chart paper, unscrew

(counter clockwise) the chart "hub" knob at the center of the chart. Gently lift the thermal

pen arm just enough to be able to slide the paper out from beneath it. Remove the old

recording chart paper and position a new one.

Re-attach the chart "hub" knob and screw securely (by hand) against the chart. Press

and hold the Change Chart button (#3) again for approximately one (1) second and the

thermal pen arm will move back onto the chart and begin recording.

Green Light LED Status:

The green LED light (located just below the three button membrane switch) is used to

show the recorder's status:

1.) LED on steady (not flashing) and input(s) recording within chart range, indicates unit

is recording normally.

2.) LED on steady (not flashing) and pen arm above outermost graduation and not

moving, indicates recorder is in Change Chart mode. Press and release Change Chart

button to return to normal recording mode.

3.) LED flashing rapidly and one or both inputs recording at outermost or innermost

graduation indicates a sensor break. Check or replace sensor(s). If sensor(s) are ok,

make sure process temperature is within configured range of recorder.

4.) LED flashing slowly (.8 seconds ON / .8 seconds OFF) indicates recorder is in

Set-Up mode. Refer to section CONFIGURING THE RECORDER.

5.) LED is Off indicates that there is no power to the recorder. Check A/C power to the

recorder.