6010 Series Operations Manual

Rev E 2/14/2012

Page 35 of 68

Operation of the Heatless Dryer Package (DRYR301)

The flow meter shows the flow rate of dry air entering the chamber. It is adjustable by

the knob at the bottom of the meter. Maximum performance is obtained at 300 SCFH

(or 5 CFM). Operating the unit above or below this purge rate may decrease

performance!! A pressure regulator is installed between the tower dryers and glass flow

meter to limit the pressure into the flow meter. The regulator should be set at a

maximum of 100psi.

1. Open the chamber door and set the humidity set point to 0%

2. Turn the flow meter adjustment knob OFF

3. Adjust the pressure regulator knob so the pressure regulator dial indicates 90 psig

maximum. (Pull up on the knob to adjust, push down on the knob to lock)

4. Adjust the flow meter adjustment knob so the flow is 300

SCFH.

The dryer will purge the chamber with compressed air only when dehumidification is

needed to reach the humidity set point.