6010 Series Operations Manual

Rev E 2/14/2012

Page 46 of 68

If a calibration adjustment is required for a single pen recorder,

use the left (#1) and right

(#2) arrow push buttons on the recorder to calibrate (or move) the pen’s position on the

chart to correspond to the temperature of the solution. The arrow buttons must be held

for approximately five seconds before the pen will begin to move.

For two pen recorders, you must first select the pen that you wish to calibrate. This is

done by pressing the left (#1) arrow button to select the red pen or the right (#2) arrow

button to select the blue pen. The arrow button must be held down until the green LED

light goes out. After the green LED light goes out, follow the instructions in step #3

above.

Battery Backup:

The green LED light remains a constant green color indicating that both the battery and

the main power to the unit are good. Refer to Figure 5 for the location of the green LED

indicating light. If the AC power were to fail or the battery becomes weak, then the green

LED light will begin "flashing" indicating that either you have lost the main power to the

unit or it is time to replace the battery. Having a 9 volt DC battery back-up in place, will

allow the recorder to continue to function normally for approximately 24 hours in the

event of a power failure.