SECTION 9 – PREVENTATIVE MAINTENANCE

The CARON chamber has been robustly designed to minimize performance problems.

However, regular maintenance is very important for continuous trouble free operation.

As a general rule, CARON recommends an annual calibration check of the temperature,

and humidity systems. CARON offers a full range of on-site calibration and validation

services. We also offer preventative maintenance contracts on our equipment. Contact

our service department for details at 740-373-6809 or visit us on the web at

www.caronproducts.com

.

Recommended Daily Maintenance Checks

Check the Temperature and humidity displays versus set-points.

Check for and correct any alarm condition.

Recommended Monthly Maintenance Checks

Check to ensure the drain in the bottom of the unit is draining properly.

Check front air intake filter. If it is dirty replace it with CARON part number FLTR303.

Washing the filter will result in poor performance.

Recommended Annual Maintenance Checks

Disinfect all interior surfaces with a general purpose laboratory cleaning agent.

Perform a complete calibration of the temperature and humidity systems.

A full validation is recommended for GMP facilities each time a unit is installed,

moved or undergoes significant repair. Contact CARON’s service department to

schedule on-site validation.