Operation of the Defrost System (DEFR301)

A Defrost System is used to maintain operational temperature below 5ºC. When the

chamber temperature is below 4.5ºC, the defrost system is enabled automatically.

When the chamber temperature is above 4.5ºC (except during a defrost cycle), the

defrost system is disabled automatically.

When the defrost system is enabled (below 4.5ºC), a defrost cycle will occur once every

12 hours and lasts for 20 minutes. During the defrost cycle, the chamber temperature

will increase several degrees above set point to melt any ice that has formed on the

evaporator. This temperature ‘spike’ is normal and assures long-term operation.