SECTION 8 – ALARMS

Alarm System Overview

The chamber control system is equipped with an alarm system that constantly monitors

temperature, and humidity (on humidified models) to ensure the user is notified if the

cabinet goes into an alarm condition. Notification occurs via a RED indicator light and

an buzzer. Each alarm condition has been factory programmed to minimize nuisance

alarms while maximizing warning time. The following table contains the alarm conditions

being checked, the factory default alarm range, the amount of time an alarm must be

present to occur (alarm delay), and the message that will be displayed on the individual

system controller.

Alarm Description

Alarm

Deviation

Alarm Delay Alarm Message

Temp higher than Set-point +1.0

C

15 minutes

Temp Controller – AL.h1

Temp lower than Set-point -1.0

C

15 minutes

Temp Controller – AL.L1

RH higher than Set-point

+5% RH

15 minutes

RH Controller – AL.h1

RH lower than Set-point

-5% RH

15 minutes

RH Controller – AL.L1

In the event an alarm occurs, the alarm indicator will illuminate and an audible alarm will

occur. To temporarily disable the audible alarm, toggle the alarm audible enable switch

to silence. When the alarm condition is corrected both the alarm indicator and the

audible alarm will be disabled. Return the alarm switch to audible.

Changing Alarm Set-points

All alarm set-points were preset at the factory to minimize nuisance alarms that could be

created as a result of door openings. Alarm set-point defaults are shown in the alarm

table earlier in this section. However, alarm set-points can be changed based on

individual user requirements. Each of the controllers are programmed in the same

manner. Press the advance key on the control system that you are changing until either

A.LO1 or A.HI1 is displayed in green. The red displayed value is the deviation from the

set-point that will activate the alarm. Press the UP arrow to increase the deviation,

press the down arrow to decrease the deviation. Press the Infinity Key to exit.