Operation of Front Mounted 6” Recorders (RCDR307, RCDR308)

Built in 6” ink pen temperature and or humidity recorders can be purchased with CARON

chambers. The recorders are shipped installed on the outer door of the chamber from

the factory and require no installation.

Changing the chart paper:

Press and hold the “change chart” button on the recorder (#3) for approximately one

second until the pen begins to move to the left of the chart and then release the button.

Wait until the pen has completely moved off of the chart. To remove the chart paper,

unscrew (counter-clockwise) the chart “hub” knob at the center of the chart. Remove the

old chart paper and position the new one so that the correct line coincides with the time

line groove on the chart plate.

Re-attach the chart “hub” knob and fasten securely against the chart. Press and hold

the “change chart” button (#3) again for approximately one second until the pen begins

to move back onto the chart and then release the button. Check to make sure that the

pen is marking on the chart paper. If it is not, then carefully adjust the pen arm to

establish contact with the paper.

Chart recorder marking system:

This type of pen consists of a self contained ink reservoir with a porous plastic stylus

which is snapped around the outer edge of the metal pen arm. A pen cap is provided to

extend the life of the ink pen during shipping or when the recording unit is not in

service. To remove the pen cap, gently lift the pen arm away from the chart paper.

Remove the black plastic pen cap to expose the fiber tip of the ink pen and gently place

the pen back onto the chart paper. Do not let the pen arm "snap" back onto the chart

paper. This will flatten the fiber tip of the pen and will no longer give you a fine line

marking on the chart paper. Place the pen cap in a safe place for future use. If the stylus

does not touch the chart, adjustment can be made by slightly bending the metal pen arm

in the center towards the chart paper. Do not use more pressure than is necessary to

create a fine line marking on the chart paper. As the pen ink supply runs out, the pen

color will become lighter. This indicates that the pen should be replaced.

Replacement of the Pen:

Recorders that are equipped with fiber tipped cartridge pens will have a cartridge that is

color coded "red" to designate pen number one and an optional cartridge that is color

coded "blue" to designate pen number two. The pen cartridge is securely fastened

to the metal pen arm using a special "U" clip tab. For ease of replacement, it is

suggested that the two screws that hold the pen arm be loosened and the pen cartridge

Advertising