MTS Series 793 User Manual

Page 321

Advertising
background image

Description

Item

Click a Rig Command button to move that channel to its specified
command level.

Rig Commands (Option)

Note:

Rig Commands are configured in the Channel Options
window.

Using Channel Limited Channel Control Mode to Install Specimens

To help protect specimens and fixtures, you may want to use the Station Builder application to create a
Channel Limited Channel (CLC) control mode that you can use as the Manual Command control mode.

The CLC control mode is typically set up as displacement control with a force signal as the limit input. Once
created, you can use the Station Setup tool to set limits (typically force) to control the amount of force that
can be applied during specimen installation.

When using Manual Command set to CLC, as the force signal value enters the user-defined CLC bandwidth,
channel limiting starts to slow the actuator response. If a force limit is reached, the limiting channel prevents
further actuator movement, regardless of the displacement command.

For more information

For more information about CLC setup and tuning, see

Adjustments Tab - CLC Control Modes

on page 234

and

About CLC Tuning

on page 236.

How to Manually Position the Actuator to Install and Remove
Specimens

This procedure is used for installing and removing specimens.

1. If desired, set up the Scope window to display the actuator’s displacement and force signals.

2.

On the Station Manager window’s Station Controls panel toolbar, click

.

3. Set up the Manual Command window:

a) Select the channel or channel group to be manually commanded.

b) Select the desired Control Mode.

c) Select Enable Manual Command.

Warning:

Applying hydraulics can result in sudden actuator motion.

A moving actuator can injure anyone in its path.

Always clear the actuator area before applying hydraulics.

MTS Series 793 Control Software 321

Using Station Manager to Run Tests

Advertising